AP-7000 Long-term deployment multi-parameter

Aquaread

New

Volume discounts

Quantity Price
(tax excl.)
You Save
%
1+ 8 683,00 € 0%
5+ 8 509,34 € 2%
10+ 8 335,68 € 4%
15+ 8 162,02 € 6%

Add to wishlist

The Sensor measures: Multiparameter: O2, pH, etc.
Cable Length: 3m
Sensor for use in: Water Quality

The AP-7000 Aquaprobe is designed for long term deployment utilising a central cleaning system to keep the fitted sensors clean reducing the effects of bio-fouling common in extended deployments. The probe comes with all of our standard parameters including optical dissolved oxygen, it also features 6 user configurable auxiliary sockets allowing you to add multiple specialist sensors to the probe. The AP-7000 is capable of measuring up to 17 individual parameters at once.

Standard parameters on the AP-7000

    Optical DO
    Specific EC
    Absolute EC
    pH
    ORP
    Depth
    TDS
    Resistivity
    Salinity
    SSG
    Temperature

The AP-7000 features 6 Auxiliary sockets to allow you to add further sensors to the probe. The sockets are completely unrestricted so you can fit the probe with any combination of our ion selective or optical sensors

The AP-7000 can be used with a variety of cable lengths via the AquaConn Connectors. Cables available in 3, 10, 20 and 30m lengths as standard.

Central Self Cleaning System

The AP-7000 is fitted with a central self cleaning system. The brush, that sits in its home position, will rotate and wipe each of the sensor faces clean multiple times per cleaning cycle. This level of cleaning maximises the duration that the probe can be left unmanned for.

Logging Options

There are various options in terms of logging:

GPS Aquameter

The GPS Aquameter is always required with any Aquaprobe. It is used for taking spot measurements, logging data manually and for calibrating the probes sensors. Its built in GPS receiver allows it to capture the location of every dataset you record. The GPS Aquameter also features a barometric pressure sensor for auto compensation for changes in air pressure during the deployment.


GPS Aquameter Provides Pin Points in Google Earth

AquaLogger

The AquaLogger is designed to work with the AP-7000. It uses standard C cell batteries to power the system and has a large memory allowing the capture of up to 15,000 datasets. The Aqualogger also features a barometric pressure sensor for auto compensation for changes in air pressure during the deployment.


The Rugged Aqualogger

The BlackBox data converter

The Blackbox is a data converter that transforms our probes output into either SDI-12 or Modbus RS485. This allows you to connect the Aquaprobe to third party logging devices, telemetry systems or existing monitoring networks. SDI-12 is our recommended protocol. It allows the Aquaprobe to effectively go to sleep between readings to reduce power consumption and on the probe waking a clean cycle is initiated prior to taking any readings.

Standard Parameters

Dissolved Oxygen Range 0 – 500.0% / 0 – 50.00 mg/L
Resolution 0.1% / 0.01mg/L
Accuracy 0 – 200%: ± 1% of reading. 200% – 500%: ± 10%
Conductivity (EC) Range 0 – 200 mS/cm (0 – 200,000 µS/cm)
Resolution 3 Auto-range scales: 0 – 9999 µS/cm, 10.00 – 99.99 mS/cm, 100.0 – 200.0mS/cm
Accuracy ± 1% of reading or ± 1µS/cm if greater
TDS* Range 0 – 100,000 mg/L (ppm)
Resolution 2 Auto-range scales: 0 – 9999mg/L, 10.00 – 100.00g/L
Accuracy ± 1% of reading or ± 1mg/L if greater
Resistivity* Range 5 Ωcm – 1 MΩcm
Resolution 2 Auto-range scales: 5 – 9999 Ωcm, 10.0 – 1000.0 KΩcm
Accuracy ± 1% of reading or ± 1 Ωcm if greater
Salinity* Range 0 – 70 PSU / 0 – 70.00 ppt (g/Kg)
Resolution 0.01 PSU / 0.01 ppt
Accuracy ± 1% of reading or ± 0.1 unit if greater
Seawater Specific Gravity* Range 0 – 50 st
Resolution 0.1 st
Accuracy ± 1.0 st
pH Range 0 – 14 pH / ± 625mV
Resolution 0.01 pH / ± 0.1mV
Accuracy ± 0.01 pH / ± 5mV
ORP Range ± 2000mV
Resolution 0.1mV
Accuracy ± 5mV
Temperature Range -5°C – +50°C (23°F – 122°F)
Resolution 0.01° C/F
Accuracy ± 0.5° C

Optional Parameters

Depth ** Range 0 – 100m (0 – 330F)
Resolution 0.01m/F
Accuracy ± 1%
**Option on the AP-2000. Comes as standard on the AP-7000

Ion Specific Electrodes***

Ammonium Range 0 – 9,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 9,000 mg/L
Accuracy ± 10% of reading
Ammonia† Range 0 – 9,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 9,000 mg/L
Accuracy ± 10% of reading
Chloride Range 0 – 20,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 20,000 mg/L
Accuracy ± 10% of reading
Flouride Range 0 – 1,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
Accuracy ± 10% of reading
Nitrate Range 0 – 30,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 30,000 mg/L
Accuracy ± 10% of reading
Calcium Range 0 – 2,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 2,000 mg/L
Accuracy ± 10% of reading
† Ammonium electrode required. Readings calculated from Ammonium, pH and temperature electrode values.

Optical Electrodes***

Turbidity Range 0 – 3000 NTU
Resolution 2 Auto-range scales: 0.0 99.9 NTU, 100 – 3000 NTU
Repeatability ± 2% of reading
Chlorophyll Range 0 – 500 µg/L (ppb)
Resolution 0.01 µg/L
Repeatability ± 2% of reading
Phycocyanin (Freshwater Blue-Green Algae) Range 0 – 300,000 cells/mL
Resolution 1 cell/mL
Repeatability ± 2% of reading
Phycoerythrin (Marine Blue-Green Algae) Range 0 – 200,000 cells/mL
Resolution 1 cell/mL
Repeatability ± 2% of reading
Rhodamine WT Dye Range 0 – 500 µg/L (ppb)
Resolution 0.01 µg/L
Repeatability ± 5% of reading
Fluorescein Dye Range 0 – 500 µg/L (ppb)
Resolution 0.01 µg/L
Repeatability ± 5% of reading
Refined Oil Range 0 – 10,000 µg/L (ppb)
Resolution 0.1 µg/L
Repeatability ± 2% of reading
CDOM/FDOM Range 0 – 2,000mg/L (ppm)
Resolution 2 Auto-range scales:
0.00 – 99.99 mg/L, 100.0 – 2,000mg/L
Repeatability ± 10% of reading or 2ppm (whichever is greater)

***Note

In addition to the standard parameters supplied with the base Aquaprobe, the AP-2000 has 1 ISE port and 1 optical port, the AP-5000 has 4 ports that can take any combination of ISE or optical sensors and the AP-7000 has 6 ports that can also take any combination of ISE or optical sensors.

The accuracy figures quoted above represent the equipment’s capability at the calibration points at 25°C. These figures do not take into account errors introduced by variations in the accuracy of calibration solutions and errors beyond the control of the manufacturer that may be introduced by environmental conditions in the field.

30 other products in the same category: