U12-015 Registrador Datos Acero Inoxidable para Temperatura -40ºC +125ºC

Onset

New

Data Logger de Acero Inoxidable para Temperatura en Agua. Rango:  -40ºC +125ºC

350,00 €

Imposto Excluido

  • Calibração ENAC de 1 ponto
  • Sem Calibração ENAC

Registrador de Datos de Acero Inoxidable USB de 1 Canal de 12-bit para Temperatura (-40ºC a +125ºC) para aplicaciones alimenticias.
Construcción en Acero Inoxidable 316L, sumergible hasta 1.500 m.
La Batería del Data Logger es 3 años y es remplazable

Tamaño: 17.5 mm diámetro x 101.6 mm; Peso: 72 gramos

Características: 

  • Rango de temperatura de alta precisión, extendida
  • Ideal para aplicaciones de alta presión para 2200psi
  • Lectura rápida a través de la interfaz USB directa

The HOBO U12 Stainless Temperature Data Logger is ideal for food, pharmaceutical, autoclave, and other applications where high-accuracy temperature data is critical. With its broad temperature range (-40 to 125C) and food-grade stainless steel housing, the U12 Stainless Temperature Data Logger withstands process conditions from pasteurization to flash freezing and washdown. The U12 Stainless is also ideal for high-temperature or deep underwater applications.

  • Food-grade 316 Series Stainless Steel
  • High-accuracy, extended temperature range
  • Suitable for high pressure applications to 2200psi
  • Fast readout via direct USB interface

Entorno: The U12-015 Data Logger is for use in Indoor , Outdoor and Underwater environments

Medidas: The U12-015 Data Logger supports the following measurements: Temperatura

Onset Data Loggers

Onset Computer Corporation

Rango de Medida: -40° to 125°C (-40° to 257°F)

Precisión: ± 0.25°C from 0° to 50°C (± 0.45°F from 32° to 122°F), see Plot A

Resolución: 0.03°C at 25°C (0.05°F at 77°F), see Plot A

Deriva: 0.05°C/year + 0.1°C/1000 hrs above 100°C (0.09°F/year + 0.2°F/1000 hrs above 212°F)

Response time in 1 m/s (2.2 mph) airflow:
U12-015: < 10 minutes, typical to 90% 
U12-015-02: 2.25 minutes, typical to 90%

Response time in water:
U12-015: < 3.5 minutes, typical to 90% 
U12-015-02: 20 seconds, typical to 90%

Time accuracy: ± 2 minute per month at 25°C (77°F), see Plot B

Operating environment: Air, water, steam, 0 to 100% RH

Operating temperature:
Logging: -40° to 125°C (-40° to 257°F)
Launch/readout: 0° to 50°C (32° to 122°F), per USB specification

Battery: life 3 year typical use, factory replaceable

Memory: 64K bytes (43,000 12-bit measurements)

Construction: Food-grade 316 Series Stainless Steel

Peso:
U12-015: 72 g (2.5 oz)
U12-015-02: 82 g (2.9 oz)

Dimensiones: 17.5 x 101.6 mm (0.69 x 4.00 inches)

Dimensiones Sonda (U12-015-02 only): 4 x 124 mm (0.16 x 4.90 inches)

Pressure/depth rating: 2200 psi (1500 m/4900 ft) maximum

Vibration rating: Navy spec: NAVMAT P-9492 (non-probe model only)

NIST certificate: Available for additional charge; temperature range -30° to 120°C (-22° to 248°F)

The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).

 

30 other products in the same category:

Clientes que compraram este produto também compraram: